4.5 Article

A prototype biosensor-integrated image-guided surgery system

Publisher

WILEY
DOI: 10.1002/rcs.123

Keywords

sensor integration; image-guided surgery; Raman spectroscopy; cancer diagnosis; medical robotics

Categories

Ask authors/readers for more resources

Background In this study we investigated the integration of a Raman spectroscopy-based biosensor with an image-guided surgery system. Such a system would provide a surgeon with both a diagnosis of the tissue being analysed (e.g. cancer) and localization information displayed within an imaging modality of choice. This type of mutual and registered information could lead to faster diagnoses and enable more accurate tissue resections. Methods A test bed consisting of a portable Raman probe attached to a passively articulated mechanical arm was used to scan and classify objects within a phantom skull. Results The prototype system was successfully able to track the Raman probe, classify objects within the phantom skull, and display the classifications on medical imaging data within a virtual reality environment. Conclusion We discuss the implementation of the integrated system, its accuracy and improvements to the system that will enhance its usefulness and further the field of sensor-based computer-assisted surgery. Copyright 2007 John Wiley & Sons, Ltd.