4.5 Article

Inversion recovery prepared turbo spin echo sequences with reduced SAR using smooth transitions between pseudo steady states

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 57, Issue 3, Pages 631-637

Publisher

JOHN WILEY & SONS INC
DOI: 10.1002/mrm.21170

Keywords

inversion recovery; TRAPS; FSE; TSE; variable flip angles; STIR; FLAIR; hyperecho; SAR

Ask authors/readers for more resources

This study investigates the contrast behavior of 2D inversion recovery (IR) prepared turbo spin echo (TSE) sequences that use RF pulse schemes with variable low flip angles (hyperTSE) to reduce RIF power deposition. A framework of equations and calculations for adapting the sequence parameters is presented by which equivalent image contrast is achieved compared to conventional IR-TSE imaging. Although the inversion time (TI) and repetition time (TR) do not need to be changed, the echo time (TE) has to be prolonged such that the effective TE (TEeff) is preserved. Measurements in healthy volunteers confirmed this finding for IR-TSE sequences using different TIs: fluid attenuated inversion recovery (FLAIR), gray matter (GM)white matter (WM)-IR, and short tau IR (STIR). The results demonstrate that hyperTSE sequences enable high-quality IR-prepared imaging with a considerably reduced specific absorption rate (SAR).

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available