4.6 Article

Amperometric detection of methanol with a methanol dehydrogenase modified electrode sensor

Journal

JOURNAL OF ELECTROANALYTICAL CHEMISTRY
Volume 601, Issue 1-2, Pages 125-131

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/j.jelechem.2006.10.039

Keywords

enzyme electrode; amperometric biosensor; methanol dehydrogenase; pyrroloquinoline quinone; methanol detection

Ask authors/readers for more resources

An amperometric enzyme electrode was developed by immobilizing the quinoprotein methanol dehydrogenase from Methylobacterium extorquens AM1 onto a glassy carbon electrode. Ferrocene (FC), ferrocene carboxylic acid, N,N,N',N'-tetramethyl-1,4-phenylenediamine (TMPD), N,N-dimethyl-p-phenylenediamine (DMPD), phenazine methosulfate (PMeS) and Wurster blue (WB) were evaluated as electron-transfer mediators. DMPD, TMPD and Fc were found to be effective electron-transfer mediators for the enzyme with only ferrocene exhibiting a stable response in the presence of oxygen. The best response for the detection of methanol was achieved with a 100 M suspension of ferrocene in 100 mM Tris/HCl buffer pH 9.0. The sensor utilized very small quantities of enzyme, and exhibited excellent reproducibility and stability with a detection limit of 0.5 mu M (S/N= 3) and a linear range of 0.5-200 mu M. (c) 2006 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available