4.7 Article

MR compatibility of EEG scalp electrodes at 4 Tesla

Journal

JOURNAL OF MAGNETIC RESONANCE IMAGING
Volume 25, Issue 4, Pages 872-877

Publisher

JOHN WILEY & SONS INC
DOI: 10.1002/jmri.20872

Keywords

EEG electrodes; MRI; B-0 field mapping; B-1 shielding; heating; safety

Ask authors/readers for more resources

Purpose: To design and apply a method to quantitatively evaluate the MR compatibility of electroencephalographic (EEG) scalp electrodes based on pulse sequence-independent metrics. Materials and Methods: Three types of electrodes (constructed primarily of brass, silver, and conductive plastic, respectively) were tested. B-0 field distortions, B-1 shielding, and heat induction was measured in adjacent agarose and oil phantoms at 4 T. B-0 field maps were corrected for distortions caused by the measurement apparatus and passive shim heating, and projections perpendicular to the surfaces of the electrodes were fit, generating cubic coefficients representing the electrode distortion severity. Signal loss in T-2-weighted images was used to determine B-1 shielding by the electrodes. Temperature measurements were recorded during the application of a high-power pulse sequence. Results: Significantly different B-0 distortions were observed in the three types of electrodes. The B-1 shielding detected in all three electrodes is minimal for most human MRI, and no significant heating was detected in the electrodes or adjacent phantom. Conclusion: The three types of electrodes were successfully differentiated in terms of MR compatibility based on pulse sequence-independent B-0 field distortions.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available