4.5 Article Proceedings Paper

Subsurface diffuse optical tomography can localize absorber and fluorescent objects but recovered image sensitivity is nonlinear with depth

Journal

APPLIED OPTICS
Volume 46, Issue 10, Pages 1669-1678

Publisher

OPTICAL SOC AMER
DOI: 10.1364/AO.46.001669

Keywords

-

Categories

Ask authors/readers for more resources

Subsurface tomography with diffuse light has been investigated with a noncontact approach to characterize the performance of absorption and fluorescence imaging. Using both simulations and experiments, the reconstruction of local subsurface heterogeneity is demonstrated, but the recovery of target size and fluorophore concentration is not linear when changes in depth occur, whereas the mean position of the object for experimental fluorescent and absorber targets is accurate to within 0.5 and 1.45 mm when located within the first 10 mm below the surface. Improvements in the linearity of the response with depth appear to remain challenging and may ultimately limit the approach to detection rather than characterization applications. However, increases in tissue curvature and/or the addition of prior information are expected to improve the linearity of the response. The potential for this type of imaging technique to serve as a surgical guide is highlighted. (c) 2007 Optical Society of America.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available