4.5 Article

Transient two-dimensional IR spectrometer for probing nanosecond temperature-jump kinetics

Journal

REVIEW OF SCIENTIFIC INSTRUMENTS
Volume 78, Issue 6, Pages -

Publisher

AIP Publishing
DOI: 10.1063/1.2743168

Keywords

-

Ask authors/readers for more resources

We have developed a Fourier transform two-dimensional infrared (2D IR) spectrometer to probe chemical reactions and biophysical processes triggered by a nanosecond temperature jump (T jump). The technical challenges for such a spectrometer involve (1) synchronization of a nanosecond T-jump laser and femtosecond laser system, (2) overcoming the decreased signal-to-noise ratio from low repetition rate data acquisition, and (3) performing an interferometric measurement through a sample with a density and index of refraction that varies with time delay after the T jump. The first challenge was overcome by synchronizing the two lasers to a clock derived from the Ti:sapphire oscillator, leading to timing accuracy of 2 ns for delays up to 50 ms. The data collection time is reduced by using undersampling with the improved signal-to-noise ratio obtained from a balanced detection scheme with a dual stripe array detector. Transient dispersed vibrational echo and 2D IR spectroscopy are applied to N-methylacetamide and ubiquitin, as examples, and the spectral responses by a temperature elevation and by structural changes of the protein are compared. The synchronization of 2D IR spectroscopy with a nanosecond temperature jump without losing its sensitivity at a low repetition rate opens a new applicability of the nonlinear spectroscopy to probe a variety of molecular structure changes induced by a nanosecond perturbation. (c) 2007 American Institute of Physics.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available