4.6 Article

Baseline correction for Raman spectra using an improved asymmetric least squares method

Journal

ANALYTICAL METHODS
Volume 6, Issue 12, Pages 4402-4407

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/c4ay00068d

Keywords

-

Funding

  1. Chongqing Key Scientific and Technological Program Project of China [cstc2012ggC70002]
  2. National Natural Science Foundation of China [21207165]
  3. CAS [KZCX2-XB3-14]
  4. West Light Foundation of Chinese Academy of Sciences
  5. Bureau of Chongqing Human Resource and Social Security [2012004]

Ask authors/readers for more resources

Baseline shifts exist in many types of Raman spectrometers. Acquired spectra normally contain the desired signals as well as undesirable elements such as background noise. In this paper, an improved asymmetric least squares (IAsLS) method has been proposed for the baseline correction of Raman spectra. The baseline correction algorithm is initiated by the raw spectrum baseline, and this baseline can be estimated using a polynomial fitting method. For the simulated Raman spectra, the performance of the proposed algorithm was evaluated and compared with the asymmetric least squares (AsLS) method and Jiang's method. The results showed that it is improved by sixteen fold and nine fold respectively. This proposed IAsLS method was successfully applied to practical Raman spectral data and the results in the paper indicate that the baseline of Raman spectra can be automatically subtracted.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available