Effect of clinical use on the cyclic fatigue resistance of ProTaper nickel-titanium rotary instruments

Journal

JOURNAL OF ENDODONTICS
Volume 33, Issue 6, Pages 737-741

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.joen.2007.03.006

Keywords

cyclic fatigue; fatigue zone; fractographic analysis; instrument separation; nickel-titanium; overload zone; rotational bending

Funding

  1. NIDCR NIH HHS [R01 DE 014911] Funding Source: Medline

Ask authors/readers for more resources

The resistance of ProTaper (Dentsply Maillefer, Ballaigues, Switzerland) nickel-titanium rotary instruments to cyclic fatigue was examined after their initial use in straight or curved canals in vivo. These instruments were rotated freely inside a steel phantom until separation. The number of rotations before failure and the lengths of the separated fragments were compared with data derived from new instruments under the same experimental setup (n = 20). With the exception of F1 and F3, instruments previously used in curved canals were more susceptible to cyclic fatigue than those previously used in straight canals (p < 0.05). Separation occurred predominantly at the D10 to D12 level. For the F series, a negative correlation (p < 0.05) was observed between the number of rotations before failure and the file diameters at their separation levels. ProTaper F3 instruments are highly susceptible to cyclic fatigue failure and should be reused with caution irrespective of whether they are initially used for shaping straight or curved canals.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available