4.6 Article

The HyperSep SCX micro-cartridge for on-line flow injection inductively coupled plasma atomic emission spectrometric determination of trace elements in biological and environmental samples

Journal

ANALYTICAL METHODS
Volume 3, Issue 9, Pages 2108-2114

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/c1ay05350g

Keywords

-

Ask authors/readers for more resources

This work covers an investigation into the potential of developing an automated on-line column preconcentration system using the readily available micro-cartridge HyperSep SCX for fast simultaneous determination of trace Ag(I), Cd(II), Co(II), Cr(III), Cu(II), Fe(III), Mn(II), Ni(II), Pb(II) and Zn(II) ions by inductively coupled plasma atomic emission spectrometry. The proposed method was based on the quantitative sorption of target elements onto the surface of the strong cation exchange resin at pH similar to 2.0, and complete elution with 2.0 mol L-1 HCl. All factors affecting the performance of the system such as sample acidity, loading and elution flow rate, preconcentration time as well as radio frequency (RF) incident power and nebulizer gas flow rate were examined in detail. The noticeable reduced consumption of chemicals along with automatic manipulations enabled the realization of a simplified and relatively clean procedure with low detection limits in the range of 0.05 and 0.2 mu g L-1. Under the optimal conditions, the developed method provided a sampling frequency of 24 h(-1) for all 10 studied elements. The accuracy of the developed method was evaluated by analyzing certified reference materials and spiked environmental natural water samples.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available