4.4 Article

Simultaneous Determination of Dopamine and Uric Acid by Electrocatalytic Oxidation on a Carbon Paste Electrode Using Pyrogallol Red as a Mediator

Journal

ANALYTICAL LETTERS
Volume 43, Issue 12, Pages 1976-1988

Publisher

TAYLOR & FRANCIS INC
DOI: 10.1080/00032711003687062

Keywords

Carbon paste electrode; Dopamine and uric acid; Electrocatalytic; Pyrogallol red

Funding

  1. Isfahan University of Technology (IUT) Research Council and Center of Excellence in Sensor (IUT)

Ask authors/readers for more resources

A sensitive and selective electrochemical method for the simultaneous determination of dopamine (DA) and uric acid (UA) was developed using a pyrogallol red modified carbon paste electrode. Under the optimized conditions, the peak current was linearly dependent on 1.0-700.0molL-1 DA and 50.0-1000.0molL-1 UA. The detection limits for DA and UA were 0.78molL-1 and 35molL-1, respectively. Finally, this method was also examined for the determination of DA and uric acid in real samples such as drugs and urine.