4.4 Article

RP-HPLC method for the determination of tenofovir in pharmaceutical formulations and spiked human plasma

Journal

ANALYTICAL LETTERS
Volume 41, Issue 4, Pages 561-570

Publisher

TAYLOR & FRANCIS INC
DOI: 10.1080/00032710801910742

Keywords

tenofovir; human plasma; RP-HPLC

Ask authors/readers for more resources

A simple reverse-phase high-performance liquid chromatographic method for the determination of tenofovir disoproxil fumarate (TDF) in pharmaceutical formulations and human plasma samples has been developed and validated. Piroxicam (PRX) was used as an internal standard. The assay of the drug was performed on a CLC C-18 (5 mu, 25 cm x 4.6 mm i.d.) with UV detection at 259 nm. The mobile phase consisted of acetonitrile - water mixture in the ratio of 75: 25, and a flow rate of 1 ml/min was maintained. The standard curve was linear over the range of 0.2 - 10 mu g/ml (r(2) = 0.9966). Analytic parameters have been evaluated. Within-day and between-day precision as expressed by relative standard deviation was found to be less than 2%. The method has been applied successfully for the determination of TDF in spiked human plasma samples and pharmaceutical formulations.