Dual Source Ion Mobility-Mass Spectrometer for Direct Comparison of Electrospray Ionization and MALDI Collision Cross Section Measurements

Journal

ANALYTICAL CHEMISTRY
Volume 82, Issue 8, Pages 3247-3254

Publisher

AMER CHEMICAL SOC
DOI: 10.1021/ac902980r

Keywords

-

Funding

  1. Vanderbilt University College of Arts and Science
  2. Vanderbilt Institute of Chemical Biology
  3. Vanderbilt Institute of Integrative Biosystems Research and Education
  4. U.S. Defense Threat Reduction Agency [HDTRA-09-1-0013]
  5. NIH National Institute on Drug Abuse [RC2DA028981]
  6. Ionwerks Inc.

Ask authors/readers for more resources

In this report, we describe a dual ionization source ion mobility-mass spectrometer (IM-MS) instrument platform for investigations that critically compare ion mobility collision cross section (CCS) measurements obtained from different ionization methods. The instrument incorporates both matrix-assisted laser desorption ionization (MALDI) and nanoelectrospray ionization (nESI) sources. The nESI source incorporates a keyhole geometry ion funnel design which facilitates axial ion focusing, accumulation, and generation of short duration (10-30 mu s) ion pulses for use with the IM-MS. The IM-MS instrument operation is independent of which ionization source is used. This allows comparisons of collision cross section measurements to be made between both ion sources with minimal differences in the instrumental arrangement. The performance of the nESI ion source is evaluated by measuring the collision cross section values of the charge states of equine cytochrome c (z = 9-16), and values are in good agreement (<2% deviation) with those previously reported in the literature. Several charge states (z = 8-11) of cytochrome c exhibit multiple cross sectional features in the ion mobility analysis. An analysis of the tryptic peptides of cytochrome c formed by both ESI and MALDI demonstrate that, on average, +1 MALDI ions are similar in CCS to +1 ESI ions and are smaller than +2 ESI ions. The ion mobility resolving power with ESI (30-35) is comparable to that obtained using MALDI (35-40), which suggests that both sources produce sufficiently narrow ion pulses for the measurement to be predominately diffusion rather than gate pulse width limited.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available