Journal
ANALYTICAL CHEMISTRY
Volume 81, Issue 13, Pages 5426-5432Publisher
AMER CHEMICAL SOC
DOI: 10.1021/ac9005827
Keywords
-
Categories
Funding
- American Society for Engineering Education
- NIH [UO1 A1075489]
- ONR/NRL 6.2 work unit [6336]
Ask authors/readers for more resources
A microfabricated flow cytometer was used to demonstrate multiplexed detection of bacteria and toxins using fluorescent coded microspheres. Antibody-coated microspheres bound biothreat targets in a sandwich immunoassay format. The microfluidic cytometer focused the microspheres in three dimensions within the laser interrogation region using passive groove structures to surround the sample stream with sheath fluid. Optical analysis at four different wavelengths identified the coded microspheres and quantified target bound by the presence of phycoerythrin tracer. The multiplexed assays in the microflow cytometer had performance approaching that of a commercial benchtop flow cytometer. The respective limits of detection for bacteria (Escherichia coli, Listeria, and Salmonella) were found to be 10(3), 10(5), and 10(4) cfu/mL for the microflow cytometer and 10(3), 10(6), and 10(5) cfu/mL for the commercial system. limits of detection for the toxins (cholera toxin, staphylococcal enterotoxin B, and ricin) were 1.6, 0.064, and 1.6 ng/mL for the microflow cytometer and 1.6, 0.064, and 8.0 ng/mL for the commercial system.