4.6 Article

Subsurface and transcutaneous Raman Spectroscopy and mapping using concentric illumination rings and collection with a circular fiber-optic array

Journal

APPLIED SPECTROSCOPY
Volume 61, Issue 7, Pages 671-678

Publisher

SAGE PUBLICATIONS INC
DOI: 10.1366/000370207781393307

Keywords

transcutaneous; fiber-optic probe; Raman spectroscopy; tomography; mapping; photon migration

Funding

  1. NIAMS NIH HHS [P30 AR46024] Funding Source: Medline
  2. NIA NIH HHS [AG024824] Funding Source: Medline
  3. NIGMS NIH HHS [T32 GM008353, T32 GM145304] Funding Source: Medline

Ask authors/readers for more resources

Different spatial separations between an illumination ring and a bundle of 50 collection fibers focused to collect light in the center of the ring were used to investigate the recovery of subsurface Raman spectra. The depth of Raman signal recovery and the preservation of spatial information in the recovered signal were investigated using polymer blocks stacked in different geometries. The illumination rings were then combined into a single data set to increase variation in the signal. Multivariate data analysis was used to recover the Raman spectra of the subsurface component. The Raman spectrum of a Delrin target was recoverable at depths up to 22.6 mm of overlying Teflon. Spatial information was lost at approximately 6.5 mm below the Teflon surface. The same protocols were used to recover canine bone spectra transcutaneously at depths up to 5 mm below the skin's surface. The recovered bone spectra were validated by exposed bone measurements.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available