4.5 Article

Development of stable isotope dilution assays for ochratoxin A in blood samples

Journal

ANALYTICAL BIOCHEMISTRY
Volume 419, Issue 2, Pages 88-94

Publisher

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.ab.2011.08.032

Keywords

Ochratoxin A; Plasma; Blood; Erythrocytes; Immunoaffinity chromatography; QuEChERS; Stable isotope dilution assay; Quantitative NMR

Ask authors/readers for more resources

Two new stable isotope dilution assays were developed for the quantification of ochratoxin A in human blood samples for exposure studies. The methods based on two different sample extraction and cleanup procedures including liquid liquid extraction with following immunoaffinity chromatography (IA) as well as a dispersive solid-phase extraction (DSPE) method. For detection, LC-MS/MS was applied. For the first time, exact quantitation of the reference compound ochratoxin A was performed by quantitative NMR spectroscopy (qNMR). Additionally, a comparison of different blood-drawing procedures revealed no differences for heparin plasma and serum whereas citrate plasma gave significantly lower results for the mycotoxin. Limits of detection (LOD: 0.02 ng/g (IA) vs 0.03 ng/g (DSPE)), limits of quantification (LOQ: 0.07 ng/g (IA) vs 0.08 ng/g (DSPE)), relative recovery ( >= 94%), precision, and linearity indicated excellent performance of the developed methods. (C) 2011 Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available