Journal
SENSORS AND ACTUATORS A-PHYSICAL
Volume 138, Issue 1, Pages 52-62Publisher
ELSEVIER SCIENCE SA
DOI: 10.1016/j.sna.2007.04.047
Keywords
SU-8; pressure sensor; Fabry-Perot; interferometer; biomedical device
Ask authors/readers for more resources
The biocompatible polymer, SU-8, was used to microfabricate an interferometric pressure sensor designed for invasive biomedical applications. Tests of the sensor in air and liquid environments show promising results as well as the limitations of SU-8 as a critical material in microdevices. The sensor consists of a polymer cap with a reflective, pressure-sensing diaphragm mounted onto the end of a fiber optic cable. Diaphragm deflection was measured by analyzing the spectrum reflected from the Fabry-Perot interferometer formed between the diaphragm and the fiber end. The device is fast, simple, and inexpensive to manufacture. Its small dimensions (300 mu m outer diameter) reduce the risk of inflammation and infection and allow for its insertion through a catheter into small vessels and cavities. The sensor showed a linear pressure response from 0 to 125 mmHg with 1-2 mmHg resolution. The cap swelled upon immersion in a manner consistent with Fickian diffusion of water into SU-8. The interferometric displacement transducer allowed a series of measurements to characterize the drift and hysteresis of the SU-8 sensors in different environments. These results provide guidance for the design and manufacture of SU-8 microdevices. (c) 2007 Elsevier B.V. All rights reserved.