4.7 Article Proceedings Paper

Ultra-fast HPLC-ICP-MS analysis of oxaliplatin in patient urine

Journal

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 397, Issue 1, Pages 401-406

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-010-3504-3

Keywords

Oxaliplatin; HPLC; ICP-MS

Funding

  1. Austrian Science Fund [P16089-N03]
  2. Austrian Science Fund (FWF) [P16089] Funding Source: Austrian Science Fund (FWF)

Ask authors/readers for more resources

A novel method for rapid HPLC-ICP-MS analysis of oxaliplatin in human urine was developed implementing a stationary HPLC phase with a particle size of 1.8 A mu m. The method allowed a cycle time of < 1 min at a HPLC flow rate of 0.9 mL min(-1). Procedural limits of detection of 0.05 A mu g L-1 oxaliplatin (150 fg on column) were obtained. Analysis of oxaliplatin in patient urine showed that accurate quantification of the intact drug demanded for storage at -80 A degrees C and rapid measurement after thawing.