A MIP-based flow-through fluoroimmunosensor as an alternative to immunosensors for the determination of digoxin in serum samples

Journal

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 394, Issue 4, Pages 963-970

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-008-2584-9

Keywords

Digoxin; Permanent immobilisation; Molecular imprinting; Flow-through fluorosensor; Serum analysis

Ask authors/readers for more resources

This work reports a comparative study of two automated flow-through fluorosensors for the determination of digoxin in serum samples: an immunosensor with an anti-digoxin polyclonal antibody as the reactive phase permanently immobilised on controlled-pore glass and a sensor with a selective reaction system based on a methacrylic molecularly imprinted polymer (MIP) synthesised by bulk polymerisation. The variables affecting the sensitivity and dynamic range of the sensors (e.g. the carrier and elution solutions, flow rates, pH and reagent concentrations) were optimized, and the binding characteristics of their reactive phases were compared in a competitive fluorescent assay. Digoxin was reproducibly determined by both sensors at the milligram per litre level (detection limit = 1.20 x 10(-3) mg L-1 and RSD = 4-7% for the immunosensor; detection limit = 1.7 x 10(-5) mg L-1 and RSD = 1-2% for the MIP sensor). No cross-reactivity with digoxin-related compounds was seen for either sensor at a digoxin/interferent ratio of 1:100. The lifetime of the immunosensor was about 50 immunoassays; its shelf life, when unused, is about 3 months. The lifetime of the MIP sensor was over 18 months. Both sensors were used to determine the digoxin concentration of human serum samples with satisfactory results.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available