4.7 Article

Use of polyclonal antibodies to ochratoxin A with a quartz-crystal microbalance for developing real-time mycotoxin piezoelectric immunosensors

Journal

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 394, Issue 2, Pages 575-582

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-009-2736-6

Keywords

Immunosensor; Quartz crystal microbalance; Piezoelectric immunosensor; Ochratoxin A; Gold nanoparticles; Voltammetry

Funding

  1. Aragon Government (Science, Technology and University Department) [PM 027/2007]
  2. ACP Company (Aragonesa de Componentes Pasivos S. A., Tarazona, Spain)

Ask authors/readers for more resources

A piezoelectric immunosensor was tested for ochratoxin A (OTA) mycotoxin detection through the immobilization of OTA-bovine serum albumin (OTA-BSA) conjugate on gold-coated quartz crystals (AT-cut/5 MHz). Immunoassays were performed in a flow-injection system through frequency decreases in a quartz-crystal microbalance (QCM) because of a mass increasing during immunoreaction with anti-OTA antibodies. Three immobilization procedures for OTA-BSA (direct adsorption and covalent attachment to two alkane thiol self-assembled monolayers) were characterized with QCM in real time. Covalent attachment of the OTA-BSA conjugates through gold nanoparticles was also tested for amplifying the signal. Binding of the excess of antibodies to the immobilized OTA in an indirect competitive analysis decreased linearly the resonant frequency in the range of the OTA concentration from 10 to 128 ng/mL, with a detection limit of 8 ng/mL (signal/noise ratio of 3). A pepsin 2 mg/mL (pH=2.1) solution was used to release antigen antibody complexes, regenerating the biorecognition surface.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available