4.7 Article

High-resolution ICP-MS determination of Ti, V, Cr, Co, Ni, and Mo in human blood and urine of patients implanted with a hip or knee prosthesis

Journal

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 391, Issue 7, Pages 2583-2589

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-008-2188-4

Keywords

HR-ICP-MS; metals release; hip and knee articular prosthesis; blood and urine samples

Ask authors/readers for more resources

The main components (Ti, V, Cr, Co, Ni, and Mo) of metallic alloys currently used in hip and knee articular prostheses have been simultaneously determined in human whole blood and urine of implanted people by a (HR)-ICP-MS method previously developed in our laboratory. The determination of those elements has been carried out in patients with knee and hip prosthesis and in a group of pre-operation patients without any metallic device in their bodies, used as controls, demonstrating the usefulness of this technique to perform multielement analysis at ppt levels in complex matrices. The concentrations of V, Cr, Co, Ni, and Mo in urine and blood of implanted people turned out to be very similar to those obtained in control patients. However, raised Ti levels could be found both in urine and blood of patients with articular prostheses made or coated with a titanium alloy (Ti6Al4V).