4.7 Article Proceedings Paper

A new ICP-TOFMS. Measurement and readout of mass spectra with 30 mu s time resolution, applied to in-torch LA-ICP-MS

Journal

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 391, Issue 4, Pages 1211-1220

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-008-1869-3

Keywords

in-torch; laser ablation; ICP-TOFMS; brass; microsecond resolution

Ask authors/readers for more resources

In-torch LA-ICP-MS was implemented into an in-house-built ICP-TOFMS system. The fast data acquisition capabilities of the new configuration allowed simultaneous multi-element measurement and readout of in-torch LA-ICP-MS signals with 30 mu s time resolution. The measurements confirmed previously observed fine structures of in-torch generated signals and provided new insights in the dynamic processes in the plasma on a microsecond time scale. The new setup is described in detail and first figures of merit are given.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available