4.6 Article

Capacitive micromachined ultrasonic transducers for chemical detection in nitrogen

Journal

APPLIED PHYSICS LETTERS
Volume 91, Issue 9, Pages -

Publisher

AMER INST PHYSICS
DOI: 10.1063/1.2776348

Keywords

-

Ask authors/readers for more resources

The authors present the prototype of a chemical sensor using a capacitive micromachined ultrasonic transducer array. Each element in the array consists of a large number of resonating membranes connected in parallel. A five-channel oscillator circuit operates at the resonant frequency around 6 MHz in this prototype. The surface of the elements in the array is coated by polymers such as polyallylamine hydrochloride, polyethylene glycol, and polyvinyl alcohol to detect different chemicals. By measuring shift in oscillation frequencies due to the mass-loading effect, analytes, e.g., water and isopropanol, with concentrations around 20 ppbv (parts per 10(9) by volume) range can be detected. (C) 2007 American Institute of Physics.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available