4.7 Article

A rapid, acetonitrile-free, HPLC method for determination of melamine in infant formula

Journal

ANALYTICA CHIMICA ACTA
Volume 665, Issue 2, Pages 227-230

Publisher

ELSEVIER
DOI: 10.1016/j.aca.2010.03.037

Keywords

Melamine; High performance liquid chromatography (HPLC); Infant formula; Validation; Acetonitrile-free

Funding

  1. Seton Hall University, Department of Chemistry and Biochemistry

Ask authors/readers for more resources

A simple, precise, accurate and validated, acetonitrile-free, reverse phase high performance liquid chromatography (HPLC) method is developed for the determination of melamine in dry and liquid infant formula. The separation is performed on a Kromasil C18 column (150 mm x 3.2 mm ID., 5 mu m particle size) at room temperature. The mobile phase (0.1% TFA/methanol 90:10) is pumped at a flow rate of 0.3 mL min(-1) with detection at 240 nm. Melamine elutes at 3.7 min. A linear response (r > 0.999) is observed for samples ranging from 1.0 to 80 mu g mL(-1). The method provides recoveries of 97.2-101.2% in the concentration range of 5-40 mu g mL(-1), intra- and inter-day variation in <1.0% R.S.D. The limit of detection (LOD) and limit of quantification (LOQ) values are 0.1 mu g mL(-1) and 0.2 mu s mL(-1), respectively. (C) 2010 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.