4.7 Article Proceedings Paper

Counting of single prion particles bound to a capture-antibody surface (surface-FIDA)

Journal

VETERINARY MICROBIOLOGY
Volume 123, Issue 4, Pages 294-304

Publisher

ELSEVIER
DOI: 10.1016/j.vetmic.2007.04.001

Keywords

prion; protemase K-free diagnosis; TSE diagnosis; single. particle detection; Surface-FIDA; BSE; scrapie; CSF

Ask authors/readers for more resources

Hitherto accredited prion tests use the PK resistance of PrPSc, the pathogenic isoform of the prion protein, as a marker for the disease. Because of variations in the amount of disease-related aggregated PrP, which is not PK-resistant, these prion tests offer only limited sensitivity. Therefore, a prion detection method that does not rely on PK digestion would allow for the detection of both PK-resistant as well as PK-sensitive PrPSc. Furthermore, single particle counting is more sensitive than methods measuring an integrated signal. Our new test system is based on dual-colour fluorescence correlation spectroscopy (FCS). This method quantifies the number of protein aggregates that have been simultaneously labelled with two different antibodies using dual-colour fluorescence intensity distribution analysis (2D-FIDA). This only counts PrP aggregates, and not PrP monomers. To increase the sensitivity, PrPSc was concentrated in a two-dimensional space by immobilizing it so that the antibodies could be captured on the surface of the slide (surface-FIDA). When the surface was systematically scanned, even single prion particles were detected. Using this new technique, the sensitivity to identify samples from scrapie-infected hamster as well as BSE-infected cattle can be dramatically increased in comparison with identification using FIDA in solution. (c) 2007 Published by Elsevier B.V.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available