4.2 Article

Flow-induced platelet activation and damage accumulation in a mechanical heart valve: Numerical studies

Journal

ARTIFICIAL ORGANS
Volume 31, Issue 9, Pages 677-688

Publisher

BLACKWELL PUBLISHING
DOI: 10.1111/j.1525-1594.2007.00446.x

Keywords

mechanical heart valve; platelets; damage; activation; blood flow; numerical methods; computational fluid dynamics; thromboembolism

Ask authors/readers for more resources

A model for platelet activation based on the theory of damage, incorporating cumulative effects of stress history and past damage (senescence) was applied to a three-dimensional (3-D) model of blood flow through a St. Jude Medical (SJM) bileaflet mechanical heart valve (MHV), simulating flow conditions after implantation. The calculations used unsteady Reynolds-averaged Navier-Stokes formulation with non-Newtonian blood properties. The results were used to predict platelet damage from total stress (shear, turbulent, deformation), and incorporate the contribution of repeated passages of the platelets along pertinent trajectories. Trajectories that exposed the platelets to elevated levels of stress around the MHV leaflets and led them to entrapment within the complex 3-D vortical structures in the wake of the valve significantly enhanced platelet activation. This damage accumulation model can be used to quantify the thrombogenic potential of implantable cardiovascular devices, and indicate the problem areas of the device for improving their designs.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available