4.5 Article

Contrast agent concentration measurements affecting quantification of bolus-tracking perfusion MRI

Journal

MAGNETIC RESONANCE IN MEDICINE
Volume 58, Issue 3, Pages 544-553

Publisher

WILEY
DOI: 10.1002/mrm.21362

Keywords

bolus-tracking MRI; cerebral blood flow; contrast agent concentration; perfusion; quantification

Ask authors/readers for more resources

Measurement of the concentration of the contrast agent using dynamic susceptibility contrast MRI relies on field inhomogeneities caused by the presence of the paramagnetic agent. The usual method for calculation of the concentration from dynamic T-2(*)-weighted images is based on two key assumptions: 1) a linear relation between the change in R-2(*) and the contrast agent concentration, and 2) a negligible effect on the MR signal due to concurrent T, changes. In this study the effect of inaccuracies in these two assumptions on perfusion measurements was investigated using simulations and in vivo data. The results of the simulations provide a quantitative characterization of the magnitude of these effects for various experimental conditions (e.g., when a 1-sec TR is used with TE = 20 ms, the T, effects can introduce up to 40% cerebral blood flow underestimation depending on the flip angle). These findings can be used as a guide to estimate the errors in specific practical implementaions, as well as to optimize the sequence parameters to minimize their effect. In summary, this study shows that the arterial input function measurement should be corrected for nonlinear R-2* effects and that care should be taken in the study design to avoid introducing significant T, effects in perfusion quantification.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available