4.6 Article Proceedings Paper

Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis

Journal

BRITISH JOURNAL OF ANAESTHESIA
Volume 99, Issue 3, Pages 337-342

Publisher

ELSEVIER SCI LTD
DOI: 10.1093/bja/aem177

Keywords

heart, cardiac output; measurement techniques, thermodilution; monitoring, arterial pressure; monitoring, cardiopulmonary; monitoring, intensive care

Categories

Ask authors/readers for more resources

Background. Recently, continuous monitoring of cardiac output (CO) based on pulse contour analysis (Vigileo((R))) has been introduced into practice. In this clinical study, we evaluated the accuracy of this system by comparing it with the transpulmonary thermodilution technique (TPID) in septic patients. Methods. We studied 24 mechanically ventilated patients with septic shock (16 male, 8 female, age 26-77 yr) receiving treatment with norepinephrine who for clinical indication underwent haemodynamic monitoring by the transpulmonary thermodilution technique using a PiCCO((R)) plus system (Pulsion Medical Systems, Munich, Germany). In parallel, arterial pulse contour was applied using the femoral arterial pressure curve (FloTrac((R)) pressure sensor, Vigileo((R)) monitor, Edwards Lifesciences, Irvine, USA). After baseline measurement, mean arterial pressure was elevated by increasing norepinephrine dosage, and CO was measured again before mean arterial pressure was reduced back to baseline levels. Fluid status and ventilator settings remained unchanged throughout. At each time point, CO by transpulmonary thermodilution was calculated from three central venous bolus injections of 15 ml of saline (< 8 degrees C). Linear regression and the Bland-Altman method were used for statistical analysis. Results. Overall, CO was 6.7 (SD 1.8) (3.2- 10.1) litre min(-1) for CO(TPID) and 6.2 (2.4) (3.0-17.6) litre min(-1) for CO(Vigileo((R))). Linear regression revealed: CO(Vigileo((R)))= 2 1.54+0.72xCO(TPID) litre min(-1), r(2) =0.26 (P < 0.0001). Mean bias between techniques [CO(TPID)-CO(Vigileo((R)))] was 0.5 litre min(-1) (SD 2.3 litre min(-1)). Correlation coefficients at the three time points were not significantly different from each other. Conclusions. Pulse contour analysis-derived CO (Vigileo((R)) system) underestimates CO(TPID) and is not as reliable as transpulmonary thermodilution in septic patients.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available