Pediatric CT Dose Reduction for Suspected Appendicitis: A Practice Quality Improvement Project Using Artificial Gaussian Noise-Part I, Computer Simulations

Journal

AMERICAN JOURNAL OF ROENTGENOLOGY
Volume 204, Issue 1, Pages W86-W94

Publisher

AMER ROENTGEN RAY SOC
DOI: 10.2214/AJR.14.12964

Keywords

appendicitis; CT; pediatrics; radiation dose reduction; size-specific dose estimate

Ask authors/readers for more resources

OBJECTIVE. The purpose of this study was to develop a departmental practice quality improvement project to systematically reduce CT doses for the evaluation of suspected pediatric appendicitis by introducing computer-generated gaussian noise. MATERIALS AND METHODS. Two hundred MDCT abdominopelvic examinations of patients younger than 20 years performed with girth-based scanning parameters for suspected appendicitis were reviewed. Two judges selected 45 examinations in which the diagnosis of appendicitis was excluded (14, appendix not visualized; 31, normal appendix visualized). Gaussian noise was introduced into axial image series, creating five additional series acquired at 25-76% of the original dose. Two readers reviewed 270 image series for appendix visualization (4-point Likert scale and arrow localization). Volume CT dose index (CTDIvol) and size-specific dose estimate (SSDE) were calculated by use of patient girth. Confidence ratings and localization accuracy were analyzed with mixed models and nonparametric bootstrap analysis at a 0.05 significance level. RESULTS. The mean baseline SSDE for the 45 patients was 16 mGy (95% CI, 12-20 mGy), and the corresponding CTDIvol was 10 mGy (95% CI, 4-16 mGy). Changes in correct appendix localization frequencies were minor. There was no substantial trend with decreasing simulated dose level (p = 0.46). Confidence ratings decreased with increasing dose reduction (p = 0.007). The average decreases were -0.27 for the 25% simulated dose (p = 0.01), -0.17 for 33% (p = 0.03), and -0.03 for 43% (p = 0.65). CONCLUSION. Pediatric abdominal MDCT can be performed with 43% of the original dose (SSDE, 7 mGy; CTDIvol, 4.3 mGy) without substantially affecting visualization of a normal appendix.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available