The number of screws, bone quality, coefficient affect acetabular cup and friction stability

Journal

MEDICAL ENGINEERING & PHYSICS
Volume 29, Issue 10, Pages 1089-1095

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2006.11.005

Keywords

acetabular cup; initial stability; screw fixation; bone quality; coefficient of friction; finite element method

Ask authors/readers for more resources

One of the major causes of loosening of cementless acetabular cup implants is insufficient initial stability. This study used three-dimensional finite element models of the pelvis and acetabular components to investigate the effects of the number of screws, bone quality, and friction coefficient of the acetabular cup on the initial stability under normal walking. A commercially available hemispherical acetabular cup with five screw holes was used as the default model. The stiffness of the pelvis and the friction coefficient of the cup were systematically varied, within a realistic range, to assess the initial stability of the acetabular cup. The simulations showed that the inserted screws provide only a localized reduction in the relative micromotion between the cup and pelvis therefore inserting several screw closed together might not be useful. Changes in the pelvic stiffness have a non-linear effect on the initial stability of acetabular cup and the subchondral bone provides good support for fixation of the cementless cup. The friction coefficient of the acetabular cup plays a limited role, comparing with the factor of bone quality, in resisting relative micromotion in the cup-pelvis interface. (C) 2006 IPEM. Published by Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available