4.5 Article

Electrical characteristics of conductive yarns and textile electrodes for medical applications

Journal

MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING
Volume 45, Issue 12, Pages 1251-1257

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s11517-007-0266-y

Keywords

textile electrodes; conductive textiles; electrode polarization; electrode impedance

Ask authors/readers for more resources

Clothing with conductive textiles for health care applications has in the last decade been of an upcoming research interest. An advantage with the technique is its suitability in distributed and home health care. The present study investigates the electrical properties of conductive yarns and textile electrodes in contact with human skin, thus representing a real ECG-registration situation. The yarn measurements showed a pure resistive characteristic proportional to the length. The electrodes made of pure stainless steel (electrode A) and 20% stainless steel/80% polyester (electrode B) showed acceptable stability of electrode potentials, the stability of A was better than that of B. The electrode made of silver plated copper (electrode C) was less stable. The electrode impedance was lower for electrodes A and B than that for electrode C. From an electrical properties point of view we recommend to use electrodes of type A to be used in intelligent textile medical applications.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available