4.7 Article

Photometric calibration of the Swift ultraviolet/optical telescope

Journal

MONTHLY NOTICES OF THE ROYAL ASTRONOMICAL SOCIETY
Volume 383, Issue 2, Pages 627-645

Publisher

OXFORD UNIV PRESS
DOI: 10.1111/j.1365-2966.2007.12563.x

Keywords

instrumentation : photometers; techniques : photometric; ultraviolet : general

Funding

  1. STFC [PP/E001173/1] Funding Source: UKRI
  2. Science and Technology Facilities Council [PP/E001173/1] Funding Source: researchfish
  3. UK Space Agency [ST/F006691/1] Funding Source: researchfish

Ask authors/readers for more resources

We present the photometric calibration of the Swift Ultraviolet/Optical Telescope (UVOT) which includes: optimum photometric and background apertures, effective area curves, colour transformations, conversion factors for count rates to flux and the photometric zero-points (which are accurate to better than 4 per cent) for each of the seven UVOT broad-band filters. The calibration was performed with observations of standard stars and standard star fields that represent a wide range of spectral star types. The calibration results include the position-dependent uniformity, and instrument response over the 1600-8000 angstrom operational range. Because the UVOT is a photon-counting instrument, we also discuss the effect of coincidence loss on the calibration results. We provide practical guidelines for using the calibration in UVOT data analysis. The results presented here supersede previous calibration results.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available