Journal
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE
Volume 21, Issue 1-2, Pages 73-86Publisher
SPRINGER
DOI: 10.1007/s10334-007-0087-x
Keywords
magnetic resonance imaging; functional imaging; echo-planar imaging; ultra high field
Ask authors/readers for more resources
Objective Single-shot echo planar imaging (EPI) acquisitions at 7T are challenging due to increased distortions, signal dropouts, RF-power requirements, and reduced T2*. This study developed and tested pulse sequence and protocol modifications required to allow high resolution EPI for whole brain functional neuroimaging. Materials and methods Using geometric distortion correction methods, modified fat saturation, and parallel imaging, we acquired high resolution single-shot gradient-echo EPI data at 7T with different spatial resolution. The BOLD sensitivity was evaluated and quantified in a breath hold experiment. Results Single-shot EPI data with isotropic resolution from 3 to 1.1 mm were acquired in human subjects. The RF-power deposition has been reduced to allow up to 22 slices per second. In addition, acoustic noise and helium boil-off have been reduced. A reduction of the fat saturation flip angle resulted in up to 20% signal gain without compromising the fat suppression quality. For the coil used, the BOLD sensitivity is highest for 2 or 1.4 mm isotropic resolution. Conclusion High resolution single-shot EPI in the whole brain can be performed at 7T with high efficiency, low signal dropout, and without major geometric distortions.