Diagnostic Capability of Spectral-Domain Optical Coherence Tomography for Glaucoma

Journal

AMERICAN JOURNAL OF OPHTHALMOLOGY
Volume 153, Issue 5, Pages 815-826

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ajo.2011.09.032

Keywords

-

Categories

Funding

  1. NATIONAL INSTITUTES OF Health, Bethesda, Maryland [R01 EY14975-01]
  2. NIDEK Co Ltd (Gamagori, Japan)
  3. Massachusetts Eye and Ear Infirmary

Ask authors/readers for more resources

PURPOSE: To determine the diagnostic capability of spectral-domain optical coherence tomography in glaucoma patients with visual field defects. DESIGN: Prospective, cross-sectional study. METHODS: SETTINGS: Participants were recruited from a university hospital clinic. STUDY POPULATION: One eye of 85 normal subjects and 61 glaucoma patients with average visual field mean deviation of -9.61 +/- 8.76 dB was selected randomly for the study. A subgroup of the glaucoma patients with early visual field defects was calculated separately. OBSERVATION PROCEDURES: Spectralis optical coherence tomography (Heidelberg Engineering, Inc) circular scans were performed to obtain peripapillary retinal nerve fiber layer (RNFL) thicknesses. The RNFL diagnostic parameters based on the normative database were used alone or in combination for identifying glaucomatous RNFL thinning. MAIN OUTCOME MEASURES: To evaluate diagnostic performance, calculations included areas under the receiver operating characteristic curve, sensitivity, specificity, positive predictive value, negative predictive value, positive likelihood ratio, and negative likelihood ratio. RESULTS: Overall RNFL thickness had the highest area under the receiver operating characteristic curve values: 0.952 for all patients and 0.895 for the early glaucoma subgroup. For all patients, the highest sensitivity (98.4%; 95% confidence interval, 96.3% to 100%) was achieved by using 2 criteria: >= 1 RNFL sectors being abnormal at the < 5% level and overall classification of borderline or outside normal limits, with specificities of 88.9% (95% confidence interval, 84.0% to 94.0%) and 87.1% (95% confidence interval, 81.6% to 92.5%), respectively, for these 2 criteria. CONCLUSIONS: Statistical parameters for evaluating the diagnostic performance of the Spectralis spectral-domain optical coherence tomography were good for early perimetric glaucoma and were excellent for moderately advanced perimetric glaucoma. (Am J Ophthalmol 2012;153:815-826. (C) 2012 by Elsevier Inc. All rights reserved.)

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available