Journal
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY
Volume 205, Issue 1, Pages -Publisher
MOSBY-ELSEVIER
DOI: 10.1016/j.ajog.2011.02.066
Keywords
fetal electrocardiogram; fetal monitoring; labor
Categories
Funding
- Center for Integration of Medicine and Innovative Technology, Boston, MA
- Obstetrix Inc, Sunrise, FL
- National Institute of Child Health and Human Development (NICHD) [1R43HD059263-01A1, 1K23HD060662-01A1]
Ask authors/readers for more resources
OBJECTIVE: We sought to evaluate the accuracy of a novel system for measuring fetal heart rate (FHR) and ST-segment changes using noninvasive electrodes on the maternal abdomen. STUDY DESIGN: Fetal electrocardiograms were recorded using abdominal sensors from 32 term laboring women who had a fetal scalp electrode (FSE) placed for a clinical indication. RESULTS: Good-quality data for FHR estimation were available in 91.2% of the FSE segments and 89.9% of the abdominal electrode segments. The root mean square error between the FHR data calculated by both methods over all processed segments was 0.36 beats per minute. ST deviation from the isoelectric point ranged from 0-14.2% of R-wave amplitude. The root mean square error between the ST change calculated by both methods averaged over all processed segments was 3.2%. CONCLUSION: FHR and ST change acquired from the maternal abdomen is highly accurate and, on average, is clinically indistinguishable from FHR and ST change calculated using FSE data.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available