Comparison of 24-hour Holter Monitoring with 14-day Novel Adhesive Patch Electrocardiographic Monitoring

Journal

AMERICAN JOURNAL OF MEDICINE
Volume 127, Issue 1, Pages -

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.amjmed.2013.10.003

Keywords

Atrial fibrillation; Cardiac arrhythmias; Electrocardiographic; Holter monitor; iRhythm; Zio Patch

Funding

  1. Clinical and Translational Science Award (National Institutes of Health/National Center for Advancing Translational Sciences) [UL1 TR000109]
  2. iRhythm Technologies Inc, San Francisco, Calif.

Ask authors/readers for more resources

BACKGROUND: Cardiac arrhythmias are remarkably common and routinely go undiagnosed because they are often transient and asymptomatic. Effective diagnosis and treatment can substantially reduce the morbidity and mortality associated with cardiac arrhythmias. The Zio Patch (iRhythm Technologies, Inc, San Francisco, Calif) is a novel, single-lead electrocardiographic (ECG), lightweight, Food and Drug Administration-cleared, continuously recording ambulatory adhesive patch monitor suitable for detecting cardiac arrhythmias in patients referred for ambulatory ECG monitoring. METHODS: A total of 146 patients referred for evaluation of cardiac arrhythmia underwent simultaneous ambulatory ECG recording with a conventional 24-hour Holter monitor and a 14-day adhesive patch monitor. The primary outcome of the study was to compare the detection arrhythmia events over total wear time for both devices. Arrhythmia events were defined as detection of any 1 of 6 arrhythmias, including supraventricular tachycardia, atrial fibrillation/flutter, pause greater than 3 seconds, atrioventricular block, ventricular tachycardia, or polymorphic ventricular tachycardia/ventricular fibrillation. McNemar's tests were used to compare the matched pairs of data from the Holter and the adhesive patch monitor. RESULTS: Over the total wear time of both devices, the adhesive patch monitor detected 96 arrhythmia events compared with 61 arrhythmia events by the Holter monitor (P < .001). CONCLUSIONS: Over the total wear time of both devices, the adhesive patch monitor detected more events than the Holter monitor. Prolonged duration monitoring for detection of arrhythmia events using single-lead, less-obtrusive, adhesive-patch monitoring platforms could replace conventional Holter monitoring in patients referred for ambulatory ECG monitoring. (C) 2014 The Authors. Published by Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available