4.3 Article

In vivo assessment of hydrocephalus shunt

Journal

ACTA NEUROLOGICA SCANDINAVICA
Volume 120, Issue 5, Pages 317-323

Publisher

WILEY
DOI: 10.1111/j.1600-0404.2009.01176.x

Keywords

hydrocephalus; shunt obstruction; in vivo testing; complications

Funding

  1. Warsaw University of Technology
  2. MRC [G9439390, G0001237, G0600986] Funding Source: UKRI
  3. Medical Research Council [G0001237, G9439390, G0600986] Funding Source: researchfish
  4. National Institute for Health Research [NF-SI-0508-10327] Funding Source: researchfish

Ask authors/readers for more resources

Objectives - Over a 3-year period, we have performed 312 tests in 197 shunted patients. The data have been analyzed retrospectively to: (1) investigate the parameters describing CSF dynamics that correlate with shunt under-drainage and (2) estimate accuracy of this method. Methods - Constant rate infusion tests into shunt prechamber were performed. Results - In 161 of the 312 infusion tests, results indicated under-draining shunts. Patients in the under-draining group had higher baseline and plateau CSF pressures, higher resistance to CSF outflow and higher levels of baseline pulse amplitude waveform. During the test, a significantly greater vasogenic waves and lower compensatory reserve was noticed in patients with blocked shunts. In 21 patients with suggestion of shunt blockage and who subsequently underwent operative revision of the shunt, reports of intraoperative shunt patency were available. Shunt blockage was confirmed intra-operatively during surgery in 19 cases. Conclusions - In vivo shunt testing is easy, safe and clinically useful, aiding decision in difficult clinical situations, where shunt malfunction is suspected but not certain. It also has satisfactory positive predictive power.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available