4.6 Article

A Novel DNA Biosensor Based on a Pencil Graphite Electrode Modified with Polypyrrole/Functionalized Multiwalled Carbon Nanotubes for Determination of 6-Mercaptopurine Anticancer Drug

Journal

INDUSTRIAL & ENGINEERING CHEMISTRY RESEARCH
Volume 54, Issue 14, Pages 3634-3639

Publisher

AMER CHEMICAL SOC
DOI: 10.1021/ie504438z

Keywords

-

Ask authors/readers for more resources

A novel and sensitive biosensor employing immobilized DNA on a pencil graphite electrode modified with polypyrrole/functionalized multiwalled carbon nanotubes for the determination of 6-mercaptopurine (6-MP) is presented. In the first step, we modified the pencil graphite surface with polypyrrole and functionalized multiwalled carbon nanotubes (MWCNT/COOH). The developed electrode was characterized by scanning electron microscopy, atomic force microscopy, reflectionabsorption infrared spectroscopy, X-ray photoelectron spectroscopy, and electrochemical impedance spectroscopy. In the other step, we used decreases in the oxidation responses of guanine and adenine as a sign of the interaction of 6-MP with salmon sperm double-stranded DNA using differential pulse voltammetry. The signal of guanine oxidation was linear with respect to the 6-MP concentration in the range of 0.2-100 mu mol L-1 with a detection limit of 0.08 mu mol L-1. The modified electrode was utilized for the determination of 6-MP in real samples.