Translation calibration of inverse-kappa goniometers in macromolecular crystallography

Journal

ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES
Volume 67, Issue -, Pages 219-228

Publisher

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S0108767311004831

Keywords

kappa goniometer; macromolecular crystallography; reorientation; calibration

Funding

  1. EU [LHSG-CT-2003-503420]

Ask authors/readers for more resources

Precise and convenient crystal reorientation is of experimental importance in macromolecular crystallography (MX). The development of multi-axis goniometers, such as the ESRF/EMBL mini-kappa, necessitates the corresponding development of calibration procedures that can be used for the setup, maintenance and troubleshooting of such devices. While traditional multi-axis goniometers require all rotation axes to intersect the unique point of the sample position, recently developed miniaturized instruments for sample reorientation in MX are not as restricted. However, the samples must always be re-centred following a change in orientation. To overcome this inconvenience and allow the use of multi-axis goniometers without the fundamental restriction of having all axes intersecting in the same point, an automatic translation correction protocol has been developed for such instruments. It requires precise information about the direction and location of the rotation axes. To measure and supply this information, a general, easy-to-perform translation calibration (TC) procedure has also been developed. The TC procedure is routinely performed on most MX beamlines at the ESRF and some results are presented for reference.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available