4.5 Article

HVAD: The ENDURANCE Supplemental Trial

Journal

JACC-HEART FAILURE
Volume 6, Issue 9, Pages 792-802

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.jchf.2018.05.012

Keywords

advanced heart failure; destination therapy; HeartWare HVAD system; LVAD; stroke

Funding

  1. Medtronic

Ask authors/readers for more resources

OBJECTIVES The aim of this study was to prospectively evaluate the impact of blood pressure management on stroke rates in patients receiving the HeartWare HVAD System. BACKGROUND The ENDURANCE trial demonstrated noninferiority of the HeartWare HVAD System versus control (HeartMate II) in patients with advanced heart failure ineligible for heart transplantation. However, stroke was more common in HVAD subjects. Post hoc analyses demonstrated increased mean arterial blood pressure as a significant independent risk factor for stroke. METHODS The ENDURANCE Supplemental Trial was a prospective, multicenter evaluation of 465 patients with advanced heart failure ineligible for transplantation, randomized 2: 1 to HVAD (n = 308) or control (n 1/4 157). The primary endpoint was the 12-month incidence of transient ischemic attack or stroke with residual deficit 24 weeks post-event. Secondary endpoints included the composite of freedom from death, disabling stroke, and need for device replacement or urgent transplantation, as well as comparisons of stroke or transient ischemic attack rates in HVAD cohorts in ENDURANCE Supplemental and ENDURANCE. RESULTS The enhanced blood pressure protocol significantly reduced mean arterial blood pressure. The primary endpoint was not achieved (14.7% with HVAD vs. 12.1% with control, noninferiority [margin 6%] p = 0.14). However, the secondary composite endpoint demonstrated superiority of HVAD (76.1%) versus control (66.9%) (p = 0.04). The incidence of stroke in HVAD subjects was reduced 24.2% in ENDURANCE Supplemental compared with ENDURANCE (p = 0.10), and hemorrhagic cerebrovascular accident was reduced by 50.5% (p = 0.02). CONCLUSIONS The ENDURANCE Supplemental Trial failed to demonstrate noninferiority of HVAD versus control regarding the pre-specified primary endpoint. However, the trial confirmed that BP management is associated with reduced stroke rates in HVAD subjects. HVAD subjects, relative to control subjects, more commonly achieved the composite endpoint (freedom from death, disabling stroke, and device replacement or urgent transplantation). (C) 2018 The Authors. Published by Elsevier on behalf of the American College of Cardiology Foundation.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available