Internal breast dosimetry in mammography: Monte Carlo validation in homogeneous and anthropomorphic breast phantoms with a clinical mammography system

Journal

MEDICAL PHYSICS
Volume 45, Issue 8, Pages 3950-3961

Publisher

WILEY
DOI: 10.1002/mp.13069

Keywords

3D-breast phantom; breast dosimetry; GEANT4; experimental measurements; dose distribution

Funding

  1. Susan G. Komen for the Cure [IIR13262248]
  2. National Cancer Institute [R01CA163746, R01CA181171]
  3. Nederlandse Organisatie voor Wetenshappelijk Onderzoek (Netherlands Organization for Scientific Research, NWO)

Ask authors/readers for more resources

PurposeTo validate Monte Carlo (MC)-based breast dosimetry estimations using both a homogeneous and a 3D anthropomorphic breast phantom under polyenergetic irradiation for internal breast dosimetry purposes. MethodsExperimental measurements were performed with a clinical digital mammography system (Mammomat Inspiration, Siemens Healthcare), using the x-ray spectrum selected by the automatic exposure control and a tube current-exposure time product of 360 mAs. A homogeneous 50% glandular breast phantom and a 3D anthropomorphic breast phantom were used to investigate the dose at different depths (range 0-4 cm with 1 cm steps) for the homogeneous case and at a depth of 2.25 cm for the anthropomorphic case. Local dose deposition was measured using thermoluminescent dosimeters (TLD), metal oxide semiconductor field-effect transistor dosimeters (MOSFET), and GafChromic films. A Geant4-based MC simulation was modified to match the clinical experimental setup. Thirty sensitive volumes (3.2 x 3.2 x 0.38 mm(3)) on the axial-phantom plane were included at each depth in the simulation to characterize the internal dose variation and compare it to the experimental TLD and MOSFET measurements. The experimental 2D dose maps obtained with the GafChromic films were compared to the simulated 2D dose distributions. ResultsDue to the energy dependence of the dosimeters and due to x-ray beam hardening, dosimeters based on these three technologies have to be calibrated at each depth of the phantom. As expected, the dose was found to decrease with increasing phantom depth, with the reduction being 93% after 4 cm for the homogeneous breast phantom. The 2D dose map showed nonuniformities in the dose distribution in the axial plane of the phantom. The mean combined standard uncertainty increased with phantom depth by up to 5.3% for TLD, 6.3% for MOSFET, and 9.6% for GafChromic film. In the case of a heterogeneous phantom, the dosimeters are able to detect local dose gradient variations. In particular, GafChromic film showed local dose variations of about 46% at the boundaries of two materials. ConclusionsResults showed a good agreement between experimental measurements (with TLD and MOSFET) and MC data for both homogeneous and anthropomorphic breast phantoms. Larger discrepancies are found when comparing the GafChromic dose values to the MC results due to the inherent less precise nature of the former. MC validations in a heterogeneous background at the level of local dose deposition and in absolute terms play a fundamental role in the development of an accurate method to estimate radiation dose. The potential introduction of a breast dosimetry model involving a nonhomogeneous glandular/adipose tissue composition makes the validation of internal dose distributions estimates crucial. (C) 2018 The Authors. Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available