4.1 Article

Clarifying the role of activated charcoal filters in preparing an anaesthetic workstation for malignant hyperthermia-susceptible patients

Journal

ANAESTHESIA AND INTENSIVE CARE
Volume 42, Issue 1, Pages 51-58

Publisher

AUSTRALIAN SOC ANAESTHETISTS
DOI: 10.1177/0310057X1404200110

Keywords

anaesthetic machine; malignant hyperthermia; activated charcoal filter

Funding

  1. Royal Melbourne Hospital Operating Suite
  2. Department of Anaesthesia and Pain Management

Ask authors/readers for more resources

Malignant hyperthermia (MH) is a life-threatening condition caused by exposure of susceptible individuals to volatile anaesthetics or suxamethonium. MH-susceptible individuals must avoid exposure to these drugs, so accurate and reproducible processes to remove residual anaesthetic agents from anaesthetic workstations are required. Activated charcoal filters (ACFs) have been used for this purpose. ACFs can reduce the time for preparing an anaesthetic workstation for MH patients. Currently, the only commercially available ACFs are the Vapor-Clean (TM) (Dynasthetics, Salt Lake City, UT, USA) filters which retail at approximately AUD$130 per set of two, both of which are to be used in a single anaesthetic. Anaesthetic workstations were saturated with anaesthetic vapours and connected to a Miran ambient air analyser (SapphIRe XL, Thermo Scientific, Waltham, MA, USA) to measure vapour concentration. Various scenarios were tested in order to determine the most economical configurations of machine flushing, component change and activated charcoal filter use. We found that placement of filters in an unprepared, saturated circuit was insufficient to safely prepare an anaesthetic workstation. Following flushing of the anaesthetic workstation with high-flow oxygen for 90 seconds, a circuit and soda lime canister change and the placement of an ACF on the inspiratory limb, we were able to safely prepare a workstation in less than three minutes. A single filter on the inspiratory limb was able to maintain a clean circuit for 12 hours, with gas flows dropped from 10 lpm to 3 lpm after 90 minutes or removal of the filter after 90 minutes if high gas flows were maintained.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available