Development of a Dedicated Rebinner with Rigid Motion Correction for the mMR PET/MR Scanner, and Validation in a Large Cohort of 11C-PIB Scans

Journal

JOURNAL OF NUCLEAR MEDICINE
Volume 59, Issue 11, Pages 1761-1767

Publisher

SOC NUCLEAR MEDICINE INC
DOI: 10.2967/jnumed.117.206375

Keywords

PET; rigid motion correction; list-mode rebinner; C-11-PIB

Funding

  1. French national 'invest for the future' programs [MR-PET ANR-11-EQPX-0026]
  2. hospital University Institut CESAME (Brain and Mental Health) [ANR-10-IBHU-0003]
  3. France Life Imaging (FLI) [ANR-11-INBS-0006]
  4. NUHS-CG CIRC
  5. SoM Aspiration Fund

Ask authors/readers for more resources

Head motion occurring during brain PET studies leads to image blurring and to bias in measured local quantities. The objective of this work was to implement a correction method for PET data acquired with the mMR synchronous PET/MR scanner. Methods: A list-mode-based motion-correction approach has been designed. The developed rebinner chronologically reads the recorded events from the Siemens list-mode file, applies the estimated geometric transformations, and frames the detected counts into sinograms. The rigid-body motion parameters were estimated from an initial dynamic reconstruction of the PET data. We then optimized the correction for C-11-Pittsburgh compound B (C-11-PIB) scans using simulated and actual data with well-controlled motion. Results: An efficient list-mode-based motion correction approach has been implemented, fully optimized, and validated using simulated and actual PET data. The average spatial resolution loss induced by inaccuracies in motion parameter estimates and by the rebinning process was estimated to correspond to a 1-mm increase in full width at half maximum with motion parameters estimated directly from the PET data with a temporal frequency of 20 s. The results show that the rebinner can be safely applied to the C-11-PIB scans, allowing almost complete removal of motion-induced artifacts. The application of the correction method to a large cohort of C-11-PIB scans led to the following observations: first, that more than 21% of the scans were affected by motion greater than 10 mm (39% for subjects with Mini-Mental State Examination scores below 20), and second, that the correction led to quantitative changes in Alzheimer-specific cortical regions of up to 30%. Conclusion: The rebinner allows accurate motion correction at a cost of minimal resolution reduction. Application of the correction to a large cohort of C-11-PIB scans confirmed the necessity of systematically correcting for motion to obtain quantitative results.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available