4.8 Article

Performance Evaluation of Series-Compensated IPT Systems for Transcutaneous Energy Transfer

Journal

IEEE TRANSACTIONS ON POWER ELECTRONICS
Volume 34, Issue 1, Pages 438-451

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TPEL.2018.2822722

Keywords

Inductive power transfer (IPT); transcutaneous energy transfer (TET)

Funding

  1. Baugarten Zurich foundation

Ask authors/readers for more resources

Today's implantable mechanical circulatory support devices, such as left ventricular assist devices, still rely on a percutaneous driveline, which is a frequent cause of severe infections and which reduces the quality of life for the patients. Inductive power transfer (IPT) is therefore a promising technology to replace the driveline and, hence, reducing the likelihood of an infection. This paper focuses on the series-series compensated IPT system and provides an in-depth comparison of two operating modes, i.e., the operation at resonance and the operation above resonance, and highlights the advantages and disadvantages with respect to the requirements set by the application at hand. In addition, the paper presents the design and the realization of a fully functional transcutaneous energy transfer (TET) implant hardware prototype, which includes the IPT front-end, the control circuit, the backup battery and its charging converter, as well as the communication electronics in a boxed volume of only 10.3 cl. The experimental verification shows that overall dc-dc efficiencies of up to 90% can be achieved for both operating modes when transmitting 25-30 W from the external battery to the implant backup battery, each having a nominal voltage of 14.8 V, using TET coils with 70 mm diameter and 10 mm coil separation distance.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available