3.8 Proceedings Paper

Multicentre knowledge sharing and planning/dose audit on flattening filter free beams for SBRT lung

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/1742-6596/573/1/012018

Keywords

-

Ask authors/readers for more resources

When implementing new technology into clinical practice, there will always be a need for large knowledge gain. The aim of this study was twofold, (I) audit the treatment planning and dose delivery of Flattening Filter Free (FFF) beam technology for Stereotactic Body Radiation Therapy (SBRT) of lung tumours across a range of treatment planning systems compared to the conventional Flatting Filter (FF) beams, (II) investigate how sharing knowledge between centres of different experience can improve plan quality. All vendor/treatment planning system (TPS) combinations investigated were able to produce acceptable treatment plans and the dose accuracy was clinically acceptable for all plans. By sharing knowledge between the different centres, the minor protocol violations (MPV) could be significantly reduced, from an average of 1.9 MPV per plan to 0.6 after such sharing of treatment planning knowledge. In particular, for the centres with less SBRT and/or volumetricmodulated arc therapy (VMAT) experience the MPV average per plan improved. All vendor/TPS combinations were also able to successfully deliver the FF and FFF SBRT VMAT plans. The plan quality and dose accuracy were found to be clinically acceptable.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

3.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available