Radiation dose reduction with frame rate conversion in X-ray fluoroscopic imaging systems with flat panel detector: basic study and clinical retrospective analysis

Journal

EUROPEAN RADIOLOGY
Volume 29, Issue 2, Pages 985-992

Publisher

SPRINGER
DOI: 10.1007/s00330-018-5620-y

Keywords

Fluoroscopy; Retrospective studies; Comparative study; X-rays; Radiation dosage

Funding

  1. JSPS (Japan Society for the Promotion of Science) KAKENHI JP Scientific Research [15K08692]
  2. Grants-in-Aid for Scientific Research [15K08692] Funding Source: KAKEN

Ask authors/readers for more resources

ObjectivesTo (a) evaluate the interpolation frames of frame rate conversion (FRC) compared with fluoroscopic frames of conventional method, and (b) compare radiation dose and fluoroscopy time between various clinical examinations without and with FRC retrospectively.MethodsThis study consisted of a basic study and a clinical retrospective analysis. The radiation dosimetry, visual assessment and measurements of contrast to noise ratio were examined. Similarity between interpolation frames and fluoroscopic frames was evaluated using normalised cross-correlation values. In the clinical retrospective analysis approved by the institutional review board, we extracted 270 examinations performed without FRC (conventional group, 12.5 pulses/s) and with FRC (FRC group, 6.25 pulses/s) from 23 May to 31 December 2016. The fluoroscopy parameters and demographics of the two groups of the clinical examinations were compared. Statistical analyses were performed with Wilcoxon signed-rank test, Brunner-Munzel test and (2) test.ResultsIn the basic study, the only significant difference was that the radiation dose of FRC was approximately half that of the conventional method in the same fluoroscopy time (p = .031). The interpolation frames of FRC were similar to the fluoroscopic frames of the conventional method. In the clinical retrospective analysis, the only significant difference was that FRC reduced the fluoroscopy dose by 48% and the total dose by 31% compared with the conventional method (p < .001). There was no significant difference in the others.ConclusionFRC significantly reduced the radiation dose without extending the fluoroscopy time and maintaining the image quality compared to the conventional method.Key Points center dot Although X-ray fluoroscopic techniques are widely used for various clinical purposes, X-ray fluoroscopic examinations have radiation risks.center dot Frame rate conversion is an image processing technique for radiation dose reduction.center dot Clinical retrospective analysis showed that FRC reduces radiation doses of patients.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available