4.7 Article

Rapid measurement of free cyanide in liquor by ion chromatography with pulsed amperometric detection

Journal

FOOD CHEMISTRY
Volume 172, Issue -, Pages 681-684

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.foodchem.2014.09.052

Keywords

Liquor; Free cyanide; Ion chromatography; Pulsed amperometric detection

Ask authors/readers for more resources

This study investigated the measurement of free cyanide in liquor by ion chromatography coupled with pulsed amperometric detection (IC-PAD). Eluent concentration, interferent evaluation and method performance were discussed. Results show that free cyanide in liquor can be rapidly determined by the optimised IC-PAD method. A sample requires only 1:100 dilution and simple filtration before being subjected to IC-PAD. The linear range is 1-5000 mu g/L with an R value of 0.9998. The detection limit is 1 mu g/L for a 25 mu L injection loop. The overall relative standard deviation (RSD) of the method is less than 5%, and the recovery range is from 98.1% to 105.0%. This study has been proven significant and may have potential applications in liquors analysis. (C) 2014 Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available