3.8 Proceedings Paper

Features of X-ray Absorption Densitometry of Large-size Objects with Variable Thickness

Publisher

IOP PUBLISHING LTD
DOI: 10.1088/1742-6596/671/1/012043

Keywords

-

Ask authors/readers for more resources

Features of formation and processing of the primary radiometric signals in the digital high-energy X-ray absorption densitometers for the homogeneous objects with variable thickness are examined. The densitometer's equation based on the polynomial approximation of the object's ray thickness dependence from its mass thickness is proposed. Guidance to select the capacity of the analog-digital converter is given. There is one example of the densitometer's equation coefficients calculation to examine the carbon, aluminum and steel wares with the mass density from 15 to 80 g/cm(2). It was shown that disagreement of the experimental and estimated values of the ray thickness for the similar mass thicknesses of the testing object is conditioned by the scattered radiation. On the high-energy digital radiography set with the X-ray source - the betatron MIB-4.5/9 the accuracy of the experimental estimation of the density was within 0.0086 g/cm(3) for the steel ware thickness from 25 to 100 mm.