4.5 Article

Three-element trap filter radiometer based on large active area silicon photodiodes

Journal

APPLIED OPTICS
Volume 55, Issue 15, Pages 3958-3965

Publisher

OPTICAL SOC AMER
DOI: 10.1364/AO.55.003958

Keywords

-

Categories

Ask authors/readers for more resources

This paper shows the opto-mechanical design of a new filter radiometer built at the Physikalisch-Technische Bundesanstalt, Germany, for the accurate determination of the thermodynamic temperature of high-temperature blackbodies. The filter radiometer is based on a three-element reflection-type trap detector that uses three large active area silicon photodiodes. Its spectral coverage and field of view are defined by a detachable narrow-band filter and a diamond-turned precision aperture, respectively. The temperature of the filter radiometer is stabilized using a water-streamed housing and is measured using a thin-film platinum thermometer placed onto the first photodiode element. The trap mount has been made as compact as possible, which, together with the large active area of the chosen photodiodes, allows a wide field of view. This work presents the design of the filter radiometer and discusses the criteria that have been considered in order for the filter radiometer to suit the application. (C) 2016 Optical Society of America

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available