3.8 Article

Whole-body positional manipulators for ocular imaging of anaesthetised mice and rats: a do-it-yourself guide

Journal

BMJ OPEN OPHTHALMOLOGY
Volume 1, Issue 1, Pages -

Publisher

BMJ PUBLISHING GROUP
DOI: 10.1136/bmjophth-2016-000008

Keywords

-

Categories

Funding

  1. Dr. Robert Pfleger Stiftung
  2. Forschungskommission der Heinrich Heine University Dusseldorf
  3. Novartis
  4. Biogen

Ask authors/readers for more resources

Background: In vivo retinal imaging of rodents has gained a growing interest in ophthalmology and neurology. The bedding of the animals with the possibility to perform adjustments in order to obtain an ideal camera-to-eye angle is challenging. Methods: We provide a guide for a cost-effective, do-it-yourself rodent holder for ocular imaging techniques. The set-up was tested and refined in over 2000 optical coherence tomography measurements of mice and rats. Results: The recommended material is very affordable, readily available and easily assembled. The holder can be adapted to both mice and rats. A custom-made mouthpiece is provided for the use of inhalant anaesthesia. The holder is highly functional and assures that the rodent's eye is the centre of rotation for adjustments in both the axial and the transverse planes with a major time benefit over unrestrained positioning of the rodents. Conclusion: We believe this guide is very useful for eye researchers focusing on in vivo retinal imaging in rodents as it significantly reduces examination times for ocular imaging.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

3.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available