4.6 Article

A Monte Carlo Analysis of Error Associated With Two-Wavelength Algorithms for Retinal Oximetry

Journal

INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE
Volume 57, Issue 14, Pages 6474-6481

Publisher

ASSOC RESEARCH VISION OPHTHALMOLOGY INC
DOI: 10.1167/iovs.16-20138

Keywords

oximetry; two-wavelength; Monte Carlo

Categories

Funding

  1. Herbert Wertheim College of Medicine
  2. Herbert and Nicole Wertheim Professorship Endowment

Ask authors/readers for more resources

PURPOSE. Two-wavelength algorithms aimed at the extrapolation of retinal vasculature optical properties are being used in the clinical setting. Although robust, this approach has some clear mathematical limitations. We have conducted an in-depth study of this methodology and report on the limits and benefit of this approach. METHODS. We used a well-tested, voxel-based Monte Carlo model of light transfer into biological tissue combined with a seven-layer model of the human fundus to create reflectance maps of retina vessels at different oxygenation levels. RESULTS. This study shows that the two-wavelength approach works remarkably well in the optimal scenario of known calibration arteries and veins. Errors as a result of choroidal pigmentation and discrepancies in vessel size can be minimized with numerical approaches. When the calibration process deviates largely from physiological values, the technique fails with large errors. CONCLUSIONS. The two-wavelength approach is convenient, easy to implement, and suitable in studies where relative rather than absolute knowledge of retinal oximetry is necessary. A robust calibration step is paramount when using this approach.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available